a) Name the common technical implementations of IMS (at least three).
- Drift tube ion mobility spectrometry (DTIMS)
- Travelling wave ion mobility spectrometry (TWIMS)
- Trapped ion mobility spectrometry (TIMS)
- Field asymmetric ion mobility spectrometry (FAIMS)
b) Which of the above IMS techniques can, in principle, be used to determine CCS values without a calibrant?
A classic drift tube with constant electric field and constant pressure, which permits to measure K0 and to calculate W directly, i.e., without needing calibrant ions for the CCS scale, while achieving 1-2 % accuracy. Thus, DTIMS is occasionally referred to as a primary method as it does not require calibration of the CCS scale. In other words, measuring lD, tD, N0, T, and E is sufficient to calculate W in DTIMS.
c) What is the best-established calibration mixture in (routine) IMS-MS?
The Agilent low concentration ESI Tune Mix (G1969-85000) can be employed in both positive-ion and negative-ion ESI mode. It has a standardized composition and its wide commercial availability made it also attractive to build an ion mobility calibration for measuring CCS values based on this ready-to-use solution. The reference values were determined via an interlaboratory evaluation, resulting in the most often used and best-defined CCS standard to date.
d) Let us consider the ion [C36H19F60N3O6P3]+. Which properties of this calibrant ion are fixed in IMS? The CCS value, K0, and/or m/z? Differentiate between cases with the same collision gas and with different collision gases.
Of course, m/z and CCS are properties of the ion. Alone m/z is constant. CCS values depend on the collision gas. K0 is determined by the actual setup in terms of collision gas used, its pressure, the electric field applied in the drift region, and of course, by the shape and size of the ion under study.