Answer 8.6

a) What is the specific advantage or unique capability of a MS-IMS-MS setup as compared to an IMS-MS/MS setup?

An MS stage in front of IMS allows for the analysis of mass-selected ions and fragment ions generated thereof by IMS. The IMS-MS/MS geometry can do tandem MS of ions separated by IMS but cannot offer preselection by m/z or fragment ion analysis by IMS.

b) There are linear and cyclic TWIMS instruments. What is the special advantage of cyclic TWIMS over TWIMS and also over DTIMS?

The IMS resolving power can finely be tuned in cyclic IMS by setting the number of cycles in the TWIMS part, i.e., elongation of the path. This achieves notably higher IMS resolving power than TWIMS, DTIMS, and even above TIMS.

c) What limitation occurs in cyclicTWIMS as the resolving power is increased? Explain.

The increasing number of cycles naturally goes along with a decrease in acquisition rate as an IMS analysis may take longer. This boost in TWIMS resolution also comes at the cost of a limited ion mobility range that can simultaneously be analyzed, as otherwise, faster ions would overtake slower ones. As discussed in the context of cyclic and multi-reflecting TOF analyzers (Chap. 6.2.11 and 6.2.12), this would prevent the unambiguous identification of the species. Thus, the observed mobility window must be selected small enough as to keep the fastest and the slowest moving ions of a batch within the period of the same single cycle.

d) How long can arrival times become in cyclic TWIMS? Up to 50 ms, up to 300 ms, or maybe up to 1500 ms?

In fact, it can take as long as 1500 ms (1.5 s)!