a) Compare the principles of operation of DTIMS and TIMS.
In DTIMS, the ions move along a gas filled tube where they are driven by a constant electric field. In TIMS, the ions are kept steady at a position of equilibrium electric field by action of that electric field. The ions align along the TIMS device and are then released by their inverted mobility.
Thus, a TIMS device effects IMS but works the opposite way to DTIMS and TWIMS. Therefore, the output of TIMS devices scales at inverse mobility, i.e., 1/K0 in units of V s cm–2.
b) Is there a constant electric field along the TIMS funnel?
No, there is an electric field gradient, i.e., the field is formed by increasingly larger voltages between subsequent electrode pairs.
c) The intial commercial version of TIMS used a single 46 mm long section. Soon after, the second generation was made of a two-stage funnel with the sections equal in length. Explain the purpose of these parts and the resulting advantage.
The front half of the device serves for ion accumulation to preserve all (or at least most) ions emerging from the ion source. The second part supplies the field gradient for alignment, then lowers the ramp to release the ions, large ones first, into the mass analyzer. So only the second part operates to separate ions by inverse mobility.
d) Decide if right or wrong?
- Reducing the pressure increases the resolving power of a TIMS device. Wrong
- Reducing the pressure expands the range of a TIMS device to larger CCS values. Right
- Narrow ranges of the inverse mobility increase the resolving power of TIMS. Right
- The resolving power of TIMS can be adjusted in the range of about 50 to 250. Right
- The upper limit of the TIMS device is around 600 Å2 for singly charged ions. Right
- Like DTIMS, TIMS does not need calibration. Wrong