a) In what regard is the field asymmetric in field asymmetric ion mobility spectrometry (FAIMS)?
The electric field is asymmetric in that the amplitude of one polarity is notably higher than that of the opposite polarity. The product of time x voltage is equal for both sections.
b) What is the „trick“ used in FAIMS to effect ion mobility separation of ions?
The separation in FAIMS is based on the fact that ions have different and notably indivdual mobilities in strong versus weak electric fields. Thus, some ions move further in the strong field phase than others, which results in some separation. It is hard to predict how an ion will behave.
c) Is FAIMS employed to measure CCS values of ions?
No. Better think of FAIMS as a zone method or mobility range filter. It can be effective in separating low-mass ions from high-mass ions, e.g., to strongly enhance protein ions while excluding the low-mass „debris“ due to impurities, solvent and/or salt cluster ions etc.
d) In what practical aspect does FAIMS differ notably from DTIMS, TWIMS, and TIMS?
The FAIMS device is very compact, in fact there is a chip variant of FAIMS. Moreover, a FAIMS device can be mounted in front of any AP interface. So it is not an integral part of an instrument.